5.Advanced User Guidebook
The description of the Channel Menu is shown as the following list:
Coupling
AC
Pass both AC and DC components of the input signal.
Block the DC component of the input signal.
Disconnect the input signal.
Inverted
Display inverted waveform.
Display original waveform.
Probe
Attenu
X10
X100
Match this to the probe attenuation factor to have an
accurate reading of vertical scale.
MeasCurr
YES
NO
If you are measuring current by probing the voltage
drop across a resistor, choose YES.
A/V (mA/V)
V/A (mV/A)
Turn the M knob to set the Amps/Volts ratio. The
range is 100mA/V ~ 1KA/V.
Amps/Volts ratio = 1/Resistor value
Volts/Amp ratio is automatically calculated.
(SDS6062(V),
SDS7072(V)
does not have
this function)
20M
Get full bandwidth.
Limit the channel bandwidth to 20MHz to reduce
display noise.
1. To set channel coupling
Taking the Channel 1 for example, the measured signal is a square wave signal
containing the direct current bias. The operation steps are shown as below:
(1) Press the CH1 MENU button to show CH1 SETUP menu.
(2) Press the H1 button, the Coupling menu will display at the screen.
(3) Press the F1 button to select the Coupling item as "DC". Both DC and AC
components of the signal are passed. The waveforms are shown as Figure
5-2.
(4) Then, press F2 button to select the Coupling item as "AC". The direct current
component of the signal is blocked. The waveforms are shown as Figure 5-3.
21